systems@systemsengineers.in

Hi 8424Portable MetersAnalytical InstrumentsHome

HI 8424Back

Microprocessor pH Meter Waterproof with Automatic Calibration & Temperature Compensation
  • Automatic
  • calibration
  • Automatic buffer recognition
  • 3 standard buffer values
  • Calibration tags on LCD
  • Low battery indicator
  • Auto shut-off
  • Back-lit display
  • HOLD function
Specification
Range : (pH) -2.00 to 16.00 pH
Range : (mV) ±699.9 mV; ±1999 mV
Range : (Temp) -20.0 to 120.0°C / -4.0 to 248.0°F
Resolution : (pH) 0.01
Resolution : (mV) 0.1 mV; 1 mV
Resolution : (Temp) 0.1°C / 0.1°F
Accuracy(@20°C) : (pH) ±0.01
Accuracy(@20°C) : (mV) ±0.2 mV; ±1 mV
Accuracy(@20°C) : (Temp) ±0.4°C / ±0.8°F
pH Calibration : automatic, 1 or 2 point with 3 memorized buffer values (pH 4.01, 7.01, 10.01)
Offset Calibration : ±1 pH
Slope Calibration : 75 to 110%
Input Impedance : 1012 Ohm
Temperature Compensation : automatic, -20 to 120°C (-4 to 248°F) or manual without temperature probe
pH Electrode : (included) HI 1230B, double junction, gel filled, PEI body, BNC conn, 1 m cable
Temperature Probe : HI 7662 with 1 m cable
Battery Type/Life : (1) 9V (included) / approx. 150 hours of continuous use; auto-off after 20 minutes of non-use or disabled (user-selectable)
Environment : 0 to 50°C (32 to 122°F); RH max 100%
Dimensions/Weight : 164 x 76 x 45 mm / 180 g
 

Overview

HI 8424 is one of the most popular pH meters on the market. By utilizing a custom microprocessor, the HI 8424 is able to perform pH, ORP and temperature measurements with a high degree of accuracy with fast response.

Calibration is automatic at 1 or 2 points with 3 memorized buffer values (pH 4.01, pH 7.01 and pH 10.01). Once the instruments has been calibrated, the buffers values used during calibration are displayed with tags on the large, back-lit LCD. This feature keeps the users informed of the current calibration and helps to avoid taking measurements that are out of range.

The mV scale auto-ranges and changes resolution from 0.1 to 1 mV when readings exceed ±699.9 mV. The extended range allows more flexibility in ORP measurements.

pH slope changes due to variation in temperature are automatically or manually compensated for.

The HOLD function freezes the reading on the LCD to be manually recorded.

Product Contents

HI 8424 is supplied with pH electrode, temperature probe, calibration and cleaning solutions, battery, protective case and instructions.

 
Home | Products | Our Clients | About Us | Contact Us Copyright © 2011 Systems Engineers.in
Web design by: Megh Technologies